TC618B Calibration: Bleeding, Running and Recording

A TC618B calibration has one problem that air never has: bubbles. A pocket of air in an oil-filled circuit compresses under pressure and expands when the pressure is released, so the reading creeps, the set point wanders, and a bench that looks stable for ten seconds drifts for the next ten minutes. Most of this TC618B calibration guide is about getting the air out and keeping it out.

TC618B calibration setup on a portable hydraulic pressure calibrator, touchscreen control panel open

What This TC618B Calibration Guide Covers

This TC618B calibration guide follows the work in the order it happens: checking the oil, bleeding the circuit, choosing points, running up and down, recording what a certificate needs, and diagnosing a point that will not hold. Specifications and ordering are on the product page.

Before you start

Check the oil level before connecting anything — mid-window is right. Low oil usually means air is being drawn in somewhere, and topping up hides the cause. Water in the oil shows as cloudiness; if the oil is cloudy or dark, change it rather than run with it.

Confirm the instrument’s span against the configuration. The 0–60 MPa bench checks instruments up to 60 MPa; a 100 MPa gauge belongs on a different source, not on this one run to its limit.

One bench with replaceable standards covers a range of instruments instead of several benches, and the accuracy claim behind it is referenced to NIST — which is the first thing an auditor asks for. For manual sources used in lower-pressure work, see our pressure test and calibration pumps.

The TC618B is built as a transportable system, so the same instrument serves the laboratory bench and the plant floor — which matters where gauges and transmitters are spread across a large site and cannot all be brought in.

Bleeding the circuit

Bleed before every job, not once a month. Connect the instrument and the reference, open the bleed point, and pressurise slowly to a low point first. Air leaves through the highest open point, so give it somewhere to go and let it go slowly — a fast pressurisation pushes the bubble into a corner instead of out.

Close the bleed point, press again, repeat. Two or three cycles is normal after a hose change; after a full drain it can take more. The test is simple: once the medium comes out of the bleed point continuously, without spitting, the circuit is full.

Fit the instrument first and the reference last. Doing it the other way round lets air back in through an open port while you are still tightening.

Choosing and running points

Read up and down. Hydraulic instruments often show hysteresis — the difference between the up and the down reading at the same point is part of what you are measuring, not noise to be averaged away. A TC618B calibration is only as good as the circuit underneath it.

Between points the bench holds its own pressure: stability reaches 0.005% F.S. on the applicable configuration, which is what lets you leave a point alone instead of nudging it.

At each point, wait for the reading to stop moving rather than for a fixed number of seconds. Above 40 MPa that wait is longer, because seals and hose are still taking up load. If the reading is still climbing after a minute, stop and look for a leak instead of waiting longer.

Never take the bench above its rated pressure to “make sure”. An over-pressure event damages the instrument and the seals, and the damage shows up later as drift you will not be able to explain.

Loop power and switch timing come from the same unit, and HART is there for the transmitters that need trimming: the pressure run and the configuration usually happen in one visit, so doing them in one place saves the second trip. Prove the loop first — 24 V applied, current reading steady — then trim.

Recording what matters

Record the up and down readings at each point, the medium temperature, and the oil condition if it is not clean. For a HART transmitter, note whether the trim was applied before or after the pressure run — a certificate that does not say is not reproducible.

Records stay in the system and export for reporting; the consistency that matters is across operators, which is what a paperless workflow buys you.

When it will not hold a point

  • Reading creeps upward at a held point — air still in the circuit, or a seal taking up load. Re-bleed before suspecting the instrument.
  • Pressure drops when the pump stops — a check valve or a fitting, almost never the gauge.
  • Handwheel goes soft near the top of the range — air in the circuit, or low oil.
  • Medium weeps at a fitting — replace the seal; do not over-tighten, a crushed seal leaks worse.
  • Up and down readings differ by more than the instrument’s tolerance — the instrument has hysteresis; record it rather than hiding it.

Maintenance

Drain and replace the oil on the interval your quality system sets, and immediately after any job where water or a contaminated medium entered the circuit. Cap the ports, coil the hoses without a sharp bend, and treat seals as consumables.

Full specifications are on the TC618B fully automatic pressure calibrator product page. For air pressure work up to 6 MPa, the TC618A portable air pressure calibrator covers the same workflow with the same control.

For a TC618B calibration on a specific instrument, send us the range and the medium — we will tell you which configuration and which points.