TC166T Manual Air Pressure Source (-0.095 to 6 MPa)

TC166T manual air pressure source

Air is the right medium for everything below about 6 MPa, and the reason is resolution. A gas column answers to the smallest valve movement, which is why this bench is quoted at 10 Pa adjustment accuracy — a hundred times finer than a liquid bench of the same size. The TC166T is built around that fact: a hand-driven pneumatic pump on a stainless-steel benchtop frame, three M20 × 1.5 ports, tool-free fittings.

No compressor and no power supply are needed: the pump is the source. Pressure is built in two stages — a lever for the bulk of the range, a screw-driven second stage for the last few hundred kilopascals — then trimmed with a fine-adjustment valve. A return-verification valve lets the operator reproduce the same point on the way down, which is where pneumatic benches usually fail: the reading drifts because a fitting is weeping, not because the pump is at fault.

TC166T manual air pressure source, alternate angle showing handwheel, lever and valve stems

Description

The TC166T is a hand-driven pneumatic pressure source for low and negative gauge pressure. It covers −0.095 to 6 MPa with 0.01% F.S stability, which puts the whole span of ordinary air instruments — vacuum gauges, 0–1.6 and 0–2.5 MPa gauges, differential transmitters — inside a single bench. Every point is set by hand, so the operator decides when the reading has settled rather than waiting on a controller.

The benchtop frame is machined from stainless steel with a powder-coated finish, and the pump, valves and ports use corrosion-resistant components. The handwheel, test ports and vent are arranged for one-operator use: build pressure with the handwheel, bleed it down with the vent, and connect the DUT through the test ports. This keeps setup time short during routine verification and protects the pressure path between calibrations.

TC166T manual air pressure source, three-quarter view with handwheel and pressure ports

Key Features

  • Hand-driven pneumatic generation — no compressor, no power supply, no services on site
  • Two-stage pump: lever pre-pressurization, screw-type secondary stage
  • 10 Pa adjustment accuracy across −0.095 to 6 MPa
  • Fine-adjustment and return-verification valves for approaching a point from either side
  • Three M20 × 1.5 ports — instrument, reference gauge and a spare at the same time
  • Dry, oil-free medium: nothing to flush out before an instrument goes back into service
  • Tool-free quick-connect fittings
  • Stainless-steel frame, powder-coated finish, 210 × 380 × 240 mm

Applications

Routine work is low-range and switch-heavy: 0–1.6 and 0–2.5 MPa gauges, differential and gauge pressure transmitters, pressure switches and low-range sensors, plus leak testing of small assemblies. Because the source is hand-driven and needs no services, it also travels — commissioning and breakdown work where the instrument cannot come to a bench.

Because the set point is held by hand, the bench also suits step-by-step verification: stop anywhere between −0.095 and 6 MPa, read the instrument, move on. Nothing is scheduled by software, so an unusual sequence costs no setup time.

Why 6 MPa Is the Practical Ceiling

Above 6 MPa, air stops being convenient. The energy stored in a compressed gas column rises with pressure, so a leak at 10 MPa is a hazard rather than an annoyance, and the volume of gas in the line makes every adjustment slow to settle. Liquid has neither problem. That is the dividing line between this bench and the hydraulic source: below 6 MPa, air gives better resolution and a cleaner instrument; above it, liquid is faster and safer.

What an Air Medium Changes

Nothing on the bench is wetted by oil, so a gauge or transmitter goes back into service without a cleaning step, and instruments with small internal volumes — pressure switches, low-range transmitters, differential sensors — are not affected by the viscosity of a liquid. The trade-off is that air leaks where oil would not. Every joint, quick-connect and test hose has to be sound; a bench that will not hold a point is almost always a fitting problem rather than a pump problem.

Where the TC166T Stops

Two cases belong elsewhere. Above 6 MPa the medium has to change, and the TC168T hydraulic power source takes over from there. If the points have to be logged and repeated without an operator, a TC618A automatic pressure calibrator runs the sequence on its own.