TC166W 250 kPa micro-pressure pump

Side view of the TC166W micro-pressure pump with handwheels and fine adjustment valve

What -50 to 250 kPa Covers

A micro-pressure pump exists because the interesting part of the low end is not the pressure, it is the resolution. 250 kPa is 2.5 bar and about 25 metres of water column, and the instruments that live down there — differential transmitters, draft and duct gauges, vacuum switches, capsule gauges on low ranges — are read in units where a coarse pump cannot land twice on the same number.

The TC166W 250 kPa micro-pressure pump covers -50 to +250 kPa in one circuit: 300 kPa of span that crosses zero, so a differential instrument can be checked on both sides of its zero in a single build instead of two set-ups with the plumbing reversed between them.

The body is 190 × 235 × 140 mm at 2.2 kg, stainless steel and aluminium, with air as the medium. That is small enough to carry to the instrument rather than bringing the instrument to a bench, which is what a micro-pressure job usually needs.

TC166W 250 kPa micro-pressure pump

Key Specifications

The figures below are for the TC166W 250 kPa micro-pressure pump as shipped. Adaptors for connection systems other than M20 × 1.5 are quoted with the order.

Pressure range-50 to +250 kPa (-0.5 to 2.5 bar)
Span300 kPa across zero
Pressure stability0.01% F.S — 30 Pa at full scale
Pressure ports2 × M20 × 1.5
Working mediumAir
ValvesFine adjustment and return verification
Dimensions190 × 235 × 140 mm
Weight2.2 kg
Power supplyNone required — hand driven

Air is the medium, so the circuit is only as rigid as its seals: every joint between the pump and the instrument is part of the measurement. That is why the fittings on the TC166W 250 kPa micro-pressure pump are quick-connect, and why it carries a built-in filter to keep the circuit clean.

The Two TC166W Ranges

Two spans share one body: -50 to +50 kPa, and the wider one sold as the TC166W 250 kPa micro-pressure pump. The pump, the fittings and the valves are the same; what changes is how wide the stability band has to be stretched.

-50 to 50 kPa-50 to 250 kPa
Span across zero100 kPa300 kPa
Stability band at full scale (0.01% F.S)10 Pa30 Pa
Pressure ports2 × M20 × 1.52 × M20 × 1.5
Working mediumAirAir

Both versions cross zero at -50 kPa. Where the whole list sits below 50 kPa, the smaller span divides the band by three — 10 Pa instead of 30 Pa — for the same pump.

Why an Air Circuit Drifts

Air is compressible, and that changes what the word stable means. In a closed air volume the pressure follows absolute temperature, so a single kelvin of ambient change moves a 250 kPa reading by roughly 1.2 kPa. Against a 30 Pa band that is forty times the tolerance, which is why temperature is not a background detail at this scale — it is the largest error term in the measurement.

The isolated pressure chamber on the TC166W 250 kPa micro-pressure pump is the answer to that. It slows the transfer of room temperature into the measuring volume, so a reading taken after the circuit has settled stays where it was put for minutes rather than seconds. It does not cancel the drift; it buys the time to record a number that is still true.

Two habits do the rest. Let the pump and the instrument sit in the same room long enough to reach the same temperature before the first build, and read after the indication has stopped moving rather than at the moment the valve is closed. On a 30 Pa band both of those matter more than the accuracy quoted for the TC166W 250 kPa micro-pressure pump.

Where 250 kPa Stops

Every instrument on the list is below 50 kPa — take the 50 kPa version; the band is 10 Pa instead of 30 for the same pump.

Anything between 50 and 250 kPa, on either side of zero — the TC166W 250 kPa micro-pressure pump covers the whole list.

Air instruments above 250 kPa — a micro-pressure pump has nothing to offer up there; the manual air pressure source runs to 6 MPa on the same principle.

Hydraulic gauges and transmitters — a different medium and a different bench; the manual hydraulic models cover 160 to 300 MPa.

The version to order is the one whose span just contains the list, because the band is set by the top of the range rather than by the pump. Where the range crosses zero and stops inside 250 kPa, that is the TC166W 250 kPa micro-pressure pump.

Get a configuration and price

TC166W 250 kPa Micro-Pressure Pump

Tell us the range you need, the number of test ports and the connection type. We reply with a matched configuration, lead time and price — usually within one business day.