Draft gauges and low-range capsule instruments are not bought in kilopascals. They are bought in inches or millimetres of water, and their spans land on round numbers in those units: a gauge reading 0 to 200 inH₂O tops out at 49.8 kPa, which is close enough that a 50 kPa span covers the whole instrument at its own full scale.
The TC166W 50 kPa micro-pressure pump covers -50 to +50 kPa — 100 kPa across zero, or ±200 inches of water. That is the working range of a draft gauge, a duct static probe, a low-range capsule gauge or a vacuum switch, on either side of zero, in a single build.
| Same range, four units | -50 kPa | +50 kPa |
|---|---|---|
| bar | -0.5 | 0.5 |
| metres of water | -5.1 | 5.1 |
| inches of water | -200 | 200 |
| psi | -7.3 | 7.3 |
The negative half is not decoration. A draft gauge reads a difference that changes sign with the fan, and a vacuum switch trips below atmospheric; both need a source that crosses zero without the plumbing being taken apart between readings — which is what the TC166W 50 kPa micro-pressure pump is for.
The body is 190 × 235 × 140 mm at 2.2 kg, stainless steel and aluminium, with air as the medium — small enough to carry to the duct or the panel rather than bringing the instrument back to a bench.
Every figure below is a delivered value for the TC166W 50 kPa micro-pressure pump, not a best case. Adaptors for other connection systems are quoted with the order.
| Pressure range | -50 to +50 kPa (-0.5 to 0.5 bar) |
| Span | 100 kPa across zero |
| Pressure stability | 0.01% F.S — 10 Pa at full scale |
| Pressure ports | 2 × M20 × 1.5 |
| Working medium | Air |
| Valves | Fine adjustment and return verification |
| Dimensions | 190 × 235 × 140 mm |
| Weight | 2.2 kg |
| Power supply | None required — hand driven |
Ten pascals is one millimetre of water column, and that is the number worth carrying into the workshop: 0.01% of a 100 kPa span. Put it against the instrument rather than against the pump. A 0–200 inH₂O gauge in the 1% accuracy class is allowed 500 Pa of error at full scale; a 10 Pa band is one fiftieth of that, which is why the source drops out of the uncertainty budget entirely at this range.
On the TC166W 50 kPa micro-pressure pump the same figure decides between the two spans. Because stability is a percentage of the span and not of the reading, 0.01% comes to 10 Pa here and to 30 Pa on the 250 kPa version. Where the list is entirely inside 50 kPa, the wider version costs three times the band for headroom that never gets used — the same pump, the same knobs, a looser number.
A narrower band does not make the medium any less elastic. Air still expands with the room, so a build left standing will wander with the temperature of the surface it stands on. Nothing in the specification prevents that; only the habit does. Bring pump and instrument to the same temperature first, then read once the indication has settled. At 10 Pa of band, that pause is worth more than any other step in the job.
Everything on the list is inside ±50 kPa — the TC166W 50 kPa micro-pressure pump is the version to order; the band is 10 Pa and the whole range is ±200 inches of water.
Instruments between 50 and 250 kPa — step up to the 250 kPa version of the same pump; the band widens to 30 Pa.
Air instruments above 250 kPa — the span, not the pump, is the obstacle; the manual air pressure source covers 6 MPa on the same air principle.
Hydraulic gauges and transmitters — wrong medium and wrong bench; the manual hydraulic models begin at 160 MPa and run to 300.
Write down the largest reading that actually appears on the certificate, in the unit the instrument is marked in, and convert once. If that number fits inside 200 inches of water, the TC166W 50 kPa micro-pressure pump covers the list with the tightest band either span can offer.
Tell us the range you need, the number of test ports and the connection type. We reply with a matched configuration, lead time and price — usually within one business day.
Range, port count and connection help us quote faster.