Tel: +86-13910814865
Email: bjyuanhuatc@gmail.com
Micro-pressure work is where hand pumps stop being practical. At 50 kPa a millimetre of piston travel moves the reading further than the tolerance allows, and holding a set point by eye while writing down a value is not a repeatable test. The TC668A-Q fully automatic pressure calibration bench removes the hand from that loop.
A servo valve drives pressure under dual-loop control — one loop on pressure, one on the rate of change — so the approach is fast and stops at the set point without overshoot. From there the bench holds that pressure while readings stabilise, then steps to the next point on its own. A batch of draft gauges, differential transmitters or low-range switches runs without an operator watching the gauge.
Range is set by the pressure module, not by the bench. Modules covering ±50, ±100 and ±250 kPa fit inside the housing or connect externally, and an internal barometric module allows gauge and absolute readings on the same instrument. That is what makes the TC668A-Q fully automatic pressure calibration bench a single purchase rather than one unit per range.
Electrical work sits on the same bench, wired independently of the pressure loop on the TC668A-Q fully automatic pressure calibration bench, so a transmitter can be checked on both sides at once: applied pressure on one screen, output current on the other. Pressure, electrical reading and the calculated error land in the same record, so a certificate does not have to be assembled from two sets of notes.
The table below lists the TC668A-Q fully automatic pressure calibration bench as delivered, in its micro-pressure configuration.
| Parameter | Micro-pressure bench — specification |
|---|---|
| Pressure range | (−50 ~ 50 / 100 / 250) kPa |
| Pressure stability | ≤ 0.003% F.S |
| Working medium | Air |
| Pressure ports | 2 × M20×1.5 (expandable) |
| Pressure standard | Configurable digital pressure module, built-in or external; an intelligent digital pressure calibrator or precision digital gauge can serve as an external reference |
| Accuracy grade | 0.01 / 0.02 / 0.05, with an internal barometric module for gauge / absolute switching |
| Control | Dual closed loop on pressure and rate of change, servo valve, automatic stabilisation |
| Pressure hold | ≥ 5 min at the set point without operator intervention |
| Display | 10-inch colour touchscreen, English and Chinese menus, 4- to 6-digit readout, live pressure and error curves |
| Communication | Wireless, WiFi, Bluetooth, LAN, RS232/485, USB |
| Dimensions | (320 × 225 × 160) mm |
| Weight | 4.5 kg |
| Protection | Overpressure alarm at 120% of range, sensor fault alarm, misconnection protection |
Twelve pressure units are selectable, including Pa, kPa, MPa, psi, bar and kgf/cm². Calibration points, cycle count and stabilisation time are editable per job, and the bench divides the range into equal steps automatically.
| Function | Range | Resolution and accuracy |
|---|---|---|
| Current measurement | (−30 ~ 30) mA | 0.1 µA; ± (0.01% RD + 0.005% FS) |
| Current output | (0 ~ 25) mA | 0.1 µA; ± (0.01% RD + 0.005% FS) |
| Voltage measurement | (−30 ~ 30) V | 0.1 mV; ± (0.01% RD + 0.005% FS) |
| Voltage measurement, high range | (−300 ~ 300) V | 0.1 µV; ± (0.01% RD + 0.005% FS) |
| Voltage output | (0 ~ 16) V | 0.1 mV; ± (0.01% RD + 0.005% FS) |
| DC output | (10 ~ 30) V | Load 50 mA; overload protection 120 mA, ± 0.5 V |
| Switch measurement | On / off | Mechanical, NPN and PNP switches, response < 10 ms |
The same electrical package is fitted to every bench in the series, so the measurement side of the TC668A-Q fully automatic pressure calibration bench does not change when the pressure range does.
| TC668A-Q bench (micro-pressure) | TC668A-Q stand (vacuum / pneumatic) | TC668A-Y bench (hydraulic) | |
|---|---|---|---|
| Range | (−50 ~ 50 / 100 / 250) kPa | (−0.095 ~ 2.5 / 4 / 7 / 10 / 15) MPa | (0 ~ 70 / 100) MPa |
| Stability | ≤ 0.003% F.S | ≤ 0.003% F.S | ≤ 0.005% F.S |
| Medium | Air | Air | Transformer oil or purified water |
| Footprint | 320 × 225 × 160 mm, 4.5 kg | 570 × 445 × 320 mm | 570 × 445 × 320 mm |
| Typical instruments | Draft gauges, differential transmitters, low-range switches, clean-room and HVAC sensors | Vacuum gauges, pneumatic transmitters and switches up to 15 MPa | Hydraulic gauges and transmitters to 100 MPa |
| Resolution limit | Kilopascal readings down to a few pascals | Pa-level at vacuum, MPa-level up high | Tens of kilopascals upward |
You need vacuum, or pneumatic work above 250 kPa — the TC668A-Q vacuum / pneumatic stand reaches −0.095 MPa and up to 15 MPa on the same control platform.
Instruments run on oil or water at high pressure — the TC668A-Y hydraulic bench goes to 100 MPa; air cannot.
The volume does not justify a bench yet — a TC166W micro-pressure pump with a digital gauge does the same range by hand, at a fraction of the cost.
Where a laboratory handles all three, the TC668A-Q fully automatic pressure calibration bench is usually the first purchase: it covers the largest number of instruments per shift, and at micro-pressure the error against tolerance is the widest. The other two follow when the range demands it.Tell us the pressure range, the accuracy grade you calibrate to and the instrument types on your bench. We reply with a matched module configuration, lead time and price — usually within one business day.