Vacuum instruments live on the wrong side of zero, and that is where testing them gets awkward. A vacuum gauge, a compound transmitter or a vacuum switch has to be checked below atmospheric pressure, at atmosphere, and often above it in the same certificate. With two separate sources the instrument is disconnected halfway through, re-sealed, and the second half of the run no longer belongs to the same test.
The TC668A-Q fully automatic pressure calibration stand removes that break. One servo-controlled source generates vacuum and positive pressure, so a compound gauge is swept through zero in a single sequence — negative points, atmospheric reference, positive points — and lands in one record. Nothing is unclamped between ranges.
Range is selected by the reference module rather than by the unit: vacuum to −0.095 MPa and positive pressure to 2.5, 4, 7, 10 or 15 MPa, with modules fitted internally or connected externally. An internal barometric module allows gauge and absolute readings on the same instrument, which is what makes the TC668A-Q fully automatic pressure calibration stand usable for both vacuum transmitters and compound gauges without a second reference.
Electrical measurement is on the same front panel, wired independently of the pressure loop, so a 4–20 mA transmitter can be read against applied pressure in one pass. Switch testing covers mechanical, NPN and PNP types — the vacuum switch that trips at a set point below atmosphere is the case this instrument was built for, and it is also the case a pressure-only bench cannot reach at all.
The table below lists the TC668A-Q fully automatic pressure calibration stand as delivered, in its vacuum / compound configuration.
| Parameter | Vacuum / pneumatic stand — specification |
|---|---|
| Pressure range | (−0.095 ~ 2.5 / 4 / 7 / 10 / 15) MPa |
| Pressure stability | ≤ 0.003% F.S |
| Working medium | Air |
| Pressure ports | 2 × M20×1.5 (expandable) |
| Pressure standard | Configurable digital pressure module, built-in or external; an intelligent digital pressure calibrator or precision digital gauge can serve as an external reference |
| Accuracy grade | 0.01 / 0.02 / 0.05, with an internal barometric module for gauge / absolute switching |
| Control | Dual closed loop on pressure and rate of change, servo valve, automatic stabilisation |
| Pressure hold | ≥ 5 min at the set point, vacuum or positive pressure |
| Display | 10-inch colour touchscreen, English and Chinese menus, 4- to 6-digit readout, live pressure and error curves |
| Communication | Wireless, WiFi, Bluetooth, LAN, RS232/485, USB |
| Dimensions | (570 × 445 × 320) mm |
| Protection | Overpressure alarm at 120% of range, sensor fault alarm, misconnection protection |
Twelve pressure units are selectable, including Pa, kPa, MPa, psi, bar and kgf/cm². Calibration points, cycle count and soak time are set per job, and the TC668A-Q fully automatic pressure calibration stand can run the whole sequence from vacuum to the top of its positive range without an operator touching a valve.
| Function | Range | Resolution and accuracy |
|---|---|---|
| Current measurement | (−30 ~ 30) mA | 0.1 µA; ± (0.01% RD + 0.005% FS) |
| Current output | (0 ~ 25) mA | 0.1 µA; ± (0.01% RD + 0.005% FS) |
| Voltage measurement | (−30 ~ 30) V | 0.1 mV; ± (0.01% RD + 0.005% FS) |
| Voltage measurement, high range | (−300 ~ 300) V | 0.1 µV; ± (0.01% RD + 0.005% FS) |
| Voltage output | (0 ~ 16) V | 0.1 mV; ± (0.01% RD + 0.005% FS) |
| DC output | (10 ~ 30) V | Load 50 mA; overload protection 120 mA, ± 0.5 V |
| Switch measurement | On / off | Mechanical, NPN and PNP switches, response < 10 ms |
Vacuum switches are commonly actuated by a 4–20 mA loop or a dry contact, so the switch channel and the current channel are usually read together. On the TC668A-Q fully automatic pressure calibration stand both update against the same pressure value, which is what makes a trip-point certificate possible from one run.
| TC668A-Q stand (vacuum / pneumatic) | TC668A-Q bench (micro-pressure) | TC668A-Y bench (hydraulic) | |
|---|---|---|---|
| Range | (−0.095 ~ 2.5 / 4 / 7 / 10 / 15) MPa | (−50 ~ 50 / 100 / 250) kPa | (0 ~ 70 / 100) MPa |
| Stability | ≤ 0.003% F.S | ≤ 0.003% F.S | ≤ 0.005% F.S |
| Medium | Air | Air | Transformer oil or purified water |
| Footprint | 570 × 445 × 320 mm | 320 × 225 × 160 mm, 4.5 kg | 570 × 445 × 320 mm |
| Crosses zero | Yes — vacuum through atmosphere to positive pressure in one run | Around zero only, within ±250 kPa | No — positive pressure only |
| Typical instruments | Vacuum gauges, compound gauges, vacuum switches, pneumatic transmitters to 15 MPa | Draft gauges, differential transmitters, clean-room and HVAC sensors | Hydraulic gauges and transmitters to 100 MPa |
Any instrument that has to be checked on both sides of atmosphere — compound gauges, vacuum transmitters, vacuum switches. This is the only bench in the series that can do it in one run.
The work tops out in kilopascals — the micro-pressure bench covers ±250 kPa at a third of the footprint and half the weight, and holds a tighter set point at that level.
Pressure is hydraulic, on oil or water — the TC668A-Y hydraulic bench reaches 100 MPa. Air cannot, and a liquid system should be tested with liquid.
Vacuum is occasional and the budget is manual — a TC166T manual air pressure source covers −0.095 to 6 MPa by hand.
Where a laboratory keeps both a vacuum line and a pneumatic line, the TC668A-Q fully automatic pressure calibration stand is normally the single replacement for the pair: one reference module set, one calibration record format, and no argument about which unit produced a certificate.
Tell us the instruments you calibrate, the pressure or vacuum range you work in, and how many test ports you need. We reply with a matched configuration, lead time and price — usually within one business day.