High pressure is generated with liquid, not gas. Compressing air to 100 MPa needs a heavy accumulator and a lot of safety margin, while a hydraulic circuit reaches the same pressure with a small volume of oil and holds it without effort. The practical consequence is that a liquid loop is stiff: connect a gauge and the pressure barely moves, so the value you set is the value you read. That is the property the TC668A-Y fully automatic pressure calibration bench is built around.
The TC668A-Y fully automatic pressure calibration bench uses that property. A servo-driven stepless valve runs dual closed loops on pressure and on rate of change, so the set point is approached quickly and then held — the operator never becomes part of the control loop. At 70 or 100 MPa a hand pump needs dozens of strokes per point and the reading drifts while the arm recovers; here the point is reached, stabilised and logged without anyone touching a valve.
Medium is a choice, not a fixed: transformer oil, or purified water where an oil film on the instrument is unacceptable — oxygen service, semiconductor, food and pharmaceutical lines. Range is set by the reference module rather than by the unit, from 0 to 70 MPa or 0 to 100 MPa, with modules fitted internally or connected externally. Twelve pressure units are selectable and two M20×1.5 ports are standard, expandable for fixtures.
Electrical work runs on the same front panel of the TC668A-Y fully automatic pressure calibration bench, independently of the pressure loop, so a 4–20 mA transmitter is read against applied pressure in one pass and HART transmitters are supported. Pointer gauges can be swept forward and in reverse for a proper hysteresis record. At this pressure level the 120% overpressure cut-out is not a formality: a 100 MPa instrument taken past range is destroyed, and the containment around it matters as much as the reading on the display.
The table below lists the TC668A-Y fully automatic pressure calibration bench as delivered.
| Parameter | Hydraulic bench — specification |
|---|---|
| Pressure range | (0 ~ 70 / 100) MPa |
| Pressure stability | ≤ 0.005% F.S |
| Working medium | Transformer oil or purified water |
| Pressure ports | 2 × M20×1.5 (expandable) |
| Pressure standard | Configurable digital pressure module, built-in or external; an intelligent digital pressure calibrator or precision digital gauge can serve as an external reference |
| Accuracy grade | 0.01 / 0.02 / 0.05, with an internal barometric module for gauge / absolute switching |
| Control | Servo-driven stepless valve, dual closed loop on pressure and rate of change |
| Pressure hold | ≥ 5 min at the set point |
| Display | 10-inch colour touchscreen, Chinese and English menus, 4- / 5- / 6-digit readout, live pressure and error curves |
| Communication | Wireless, WiFi, Bluetooth, LAN, RS232/485, USB |
| Dimensions | (570 × 445 × 320) mm |
| Protection | 120% overpressure cut-out, sensor fault alarm, misconnection protection |
Calibration points, cycle count and stabilisation time are editable per job. The TC668A-Y fully automatic pressure calibration bench runs up and down cycles without an operator present, which is what turns a batch of gauges into one run instead of a series of separate setups.
| Function | Range | Resolution and accuracy |
|---|---|---|
| Current measurement | (−30 ~ 30) mA | 0.1 µA; ± (0.01% RD + 0.005% FS) |
| Current output | (0 ~ 25) mA | 0.1 µA; ± (0.01% RD + 0.005% FS) |
| Voltage measurement | (−30 ~ 30) V | 0.1 mV; ± (0.01% RD + 0.005% FS) |
| Voltage measurement, high range | (−300 ~ 300) V | 0.1 µV; ± (0.01% RD + 0.005% FS) |
| Voltage output | (0 ~ 16) V | 0.1 mV; ± (0.01% RD + 0.005% FS) |
| DC output | (10 ~ 30) V | Load 50 mA; overload protection 120 mA, ± 0.5 V |
| Switch measurement | On / off | Mechanical, NPN and PNP switches, response < 10 ms |
Every channel is read against the same pressure value, so pressure, electrical reading and calculated error end up in one record. Records export to PC and print to certificate, with the error curve attached.
| TC668A-Y (hydraulic bench) | TC618B (portable hydraulic) | TC668A-Q stand (vacuum / pneumatic) | |
|---|---|---|---|
| Range | (0 ~ 70 / 100) MPa | (0 ~ 60) MPa | (−0.095 ~ 15) MPa |
| Stability | ≤ 0.005% F.S | 0.005% F.S | ≤ 0.003% F.S |
| Medium | Transformer oil or purified water | Transformer oil or purified water | Air |
| Form | 570 × 445 × 320 mm bench | Carrying case, battery powered | 570 × 445 × 320 mm bench |
| Best at | Batch work in a lab or on a production line | Calibrating the instrument where it is installed | Gas, vacuum and compound instruments |
| Above 60 MPa | Yes, to 100 MPa | No | No |
Every channel is read against the same pressure value, so pressure, electrical reading and calculated error end up in one record. Records export to PC and print to certificate, with the error curve attached.
Hydraulic instruments above 60 MPa, or the same test repeated all day — this bench. It is the only one in the range that reaches 100 MPa, and the only one built for unattended cycles.
The instruments are spread across a plant, up to 60 MPa — the TC618B portable hydraulic calibrator goes to the gauge instead of the gauge coming to the bench.
Gas, vacuum or compound instruments — the TC668A-Q vacuum / pneumatic stand covers −0.095 to 15 MPa in air. Oil in a gas loop is a cleaning problem nobody wants.
Hydraulic work is occasional and the budget is manual — a TC168T manual hydraulic source reaches 100 MPa by hand.
Where a laboratory verifies both hydraulic and pneumatic instruments, the two benches are usually bought as a pair: the TC668A-Y fully automatic pressure calibration bench takes everything on oil, the Q stand takes everything on air, and both write the same record format.
Once the bench is on site, the routine is the sequence itself: which modules to load, how long to dwell at each point, and how the record is produced afterwards. That is covered step by step in the TC668A-Y calibration guide.
Tell us the instruments you calibrate, the range you work in, and how many test ports you need. We reply with a matched configuration, lead time and price — usually within one business day.
Range, medium and quantity help us quote faster.